Calibration consistency: check it, fix it, and only then patch it

In short

The same particle must get the same position from all cameras and from any three of them. Check it in seconds; fix it by fitting the cameras again or by aligning the partial points. Never adopt the fix from the report alone.

Basics: Course page 8. Do it: Tracks break when the set of cameras changes.

A 3D-PTV run is only as good as the agreement between its cameras. When a particle is seen by all four cameras, four rays meet in one point. When one camera drops out, the three remaining rays meet in a slightly different point if the calibration is not perfectly consistent. The measured position jumps whenever a track changes between "4 cameras" and "3 cameras, camera m missing". Averaged over many particles the jump is a systematic offset, and a velocity (a difference of two positions) does not cancel it. It is also the cheapest direct measure of calibration quality that a finished run offers, and it needs no ground truth.

This page describes three layers, in the order to use them.

1. Check (always; seconds)

uv run python -m openptv2.calibration_consistency check <run folder>            # parameters_Run1.yaml, cal/, res/run.zarr
from openptv2.calibration_consistency import consistency_report
print(consistency_report(store, frames, cals, cpar, vpar).text())

It reports, per camera, how far the blob centres lie from the projection of the 4-camera position, the smooth part of that (the residual field), whether that field is real (a field fitted on the odd frames explains the even frames; noise does not; "n/a" when there is no second half of frames), and the systematic jump per missing camera in micrometres and in pixel footprints. The verdict:

verdict worst jump (px footprint) meaning
inconclusive any the offsets are within 3 standard errors of the noise of this sample (too few points / frames): measure on more
consistent < 0.05 nothing to do
marginal 0.05 – 0.30 small systematic offsets: re-calibrate if you can, otherwise align (layer 3)
inconsistent > 0.30 re-calibrate (layer 2)

Points whose four cameras disagree (wrong matches: ghosts, merged blobs; rms reprojection residual above max(1.5 px, 3 x median)) are left out of every number and their share is printed; on the cavity data they were 12 % and had inflated a 30 µm jump to 200 µm.

Reference points (400 and 300 frames, then the full 1072): a perfect calibration shows 0.02 px (field reproducibility 0–2 %, i.e. noise); the real recording wp2 shows 0.14 px (marginal; reproducibility ~50 %); the rendered syn1 with its real calibration error shows 0.47 px (inconsistent). With fewer than four cameras the check says so and does nothing.

2. Re-calibrate from the tracers of the run (the fix)

uv run python -m openptv2.calibration_consistency selfcal <run folder> [--out new_cal_folder]

self_calibrate_from_store fits exterior orientation, principal point and k1 k2 p1 p2 of every camera on the clean 4-camera points of the odd frames and judges the result on the even frames. The new cameras are accepted only when (1) the worst jump on held-out frames falls by at least half, and (2) matching those frames again with the new cameras does not lose 4-camera matches (min_yield, ratio >= 1.0; --min-yield). Lens terms are fitted with priors set in pixels at the sensor corner (edge_px, default 1 px). One camera's exterior is held fixed as the reference (hold_cam, default "auto": the camera that already agrees best); a held camera keeps its own errors for good and the others are moved to agree with it, which the consistency numbers cannot show (docs/self-calibration-cavity-synthetic.md, section 2). It declines when the cameras were already consistent, and when the fit does not clearly help (then the inconsistency is not a camera-parameter error).

Why the second test: the jump is measured on points that the old cameras matched, and the fit saw the same kind of points, so the jump can shrink while the calibration gets worse for the particles as a whole. That happened (below). The yield ratio is the cheapest independent signal we found.

When self-calibration helped, and when it did not

The first version of the fit lost matches and tracks on lv_multi because its priors on the lens terms were meaningless: the distortion acts on the image radius in mm (k1 ~ r^3, k2 ~ r^5), so a fixed prior number means thousands of pixels at the sensor edge, the data fix the terms near the centre, and the fit then wrecked the periphery (97 % of the matches beyond 5 mm lost). The priors are now set in pixels at the sensor corner (edge_px, 1 px). With that:

data cameras report yield with new cameras two_phase / trackcorr / fast_3d
wp2 (300 frames) one side marginal +11 % tracks >= 50 frames +11.7 %; points in tracks >= 20 frames +10.5 %
wp1 (300 frames) one side marginal +9 % tracks >= 50 frames +15.2 %; points in tracks >= 20 frames +10.9 %
wp1 / wp2, all 1072 frames one side marginal not measured tracks >= 50 frames +9.3 / +9.8 % with the old priors
lv_multi wp5 (20 frames) one side marginal +9 % good trajectories +1.4 / +5.4 / +4.9 % (were -9.7 / -1.5 / -1.6 %)
lv_multi wp4 (20 frames) one side marginal +8 % -0.5 / +5.1 / +3.8 % (were -7.4 / -1.5 / -3.2 %)
rendered syn1 (truth) one side inconsistent see the lab page velocity error -26 ... -42 % (fast_3d)
test_cavity (5 frames) opposite marginal, field not reproducible +2 % noise-level; blob noise dominates

Still true, and still the main caveat: the volume changes (docs/self-calibration-scale-bias.md); and the matching-yield acceptance test remains, because a smaller jump among points the old cameras matched is not proof of a better calibration. Match again and compare your own tracking before adopting new cameras. The CLI writes the new cameras to a new folder, so trying costs nothing.

3. Align the partial-camera points (stopgap; explicit; marked)

uv run python -m openptv2.calibration_consistency align <run folder>

align_partial_points subtracts the smooth 4-camera residual field from the blob centres of the 2- and 3-camera points and triangulates them again. 4-camera points are not touched at all (tested: zero changed), so the absolute frame and scale of the calibration stay exactly as they were. It is never run implicitly. It refuses when the cameras are already consistent, when the stored points were matched with other calibration files (a re-calibration without new matching), or when it was already applied; on success the run store gets the attribute align_partial_points.

Where it is useful in practice

Its one real advantage over layer 2 is that it does not change the gauge. Self-calibration moves the volume (see docs/self-calibration-scale-bias.md: stretches of 0.2-1 %); alignment cannot. Use it when:

  1. The validated scale and frame must stay. The calibration was checked against a dumbbell, a plate or a known length, or it is shared with other recordings or earlier campaigns, and a change of 0.3 % in scale is not acceptable. You still want the camera-switching jump out of tracks that change between 4 and 3 cameras.
  2. You cannot or may not change the cameras. A certified or published calibration, or a pipeline that takes .ori files as given, or no time to re-match.
  3. The report says marginal and the quantity you care about is sensitive to small systematic offsets between camera subsets (accelerations, small-scale statistics of positions), not just track counts.
  4. A quick, local fix on an already matched store. It takes seconds, needs no new matching and no new .ori files. Run the trackers afterwards.

When to avoid it

situation why
the report says consistent nothing to correct (it refuses by default)
the report says inconclusive the offsets are within the noise of the sample; measure on more frames first
the report says inconsistent alignment removes the jump, not the camera error: the 4-camera points stay as wrong as they are. Re-calibrate (layer 2): it was better in every measurement
you will re-calibrate anyway after layer 2 and a new matching the report reads consistent; alignment would only add risk
the legacy trackcorr is your tracker it changed nothing (error doubled: identical output) or little (real error: velocity error 0.574 → 0.552, -4 %) in our tests: it works from the images, not from the stored 3D points
few partial-camera points the gain scales with how many 3-camera points there are
most partial points lie far from the 4-camera region the field is fitted on 4-camera points; far outside them it is extrapolated (on wp2 up to 1.1 mm for the 12 % of the 3-camera points farthest away; unverified)
fewer than 4 cameras not applicable
the stored points do not come from the cameras you pass refused: it would move partial points by the calibration difference only
you need to undo it it rewrites the stored correspondences in place and records a marker; to undo, match the frames again
you also use shared_pass, light_check, shared_correction the combinations were not measured

What it did in our tests

data tracker effect
rendered syn1, real error fast_3d common-set velocity error 0.311 → 0.277 (-11 %), position error -12 %
rendered syn1, error doubled fast_3d 0.405 → 0.423 (+5 %, worse); two_phase position error -19 %; trackcorr unchanged
rendered syn1, perfect calibration any none (it refuses; forced: 2 µm correction)
real wp2 (two ranges), wp1 two_phase tracks of 50 frames or longer +3.3 %, +2.7 %, 0 % (wp1 had almost no jump in its tracks: 2-4 µm)

So expect a small gain at best on real recordings, a possible loss when the error is large, and nothing with the legacy tracker. Treat it as the gauge-preserving alternative to layer 2, not as a substitute.

What the synthetic cavity showed (truth known, cameras on both sides)

Details and tables: docs/self-calibration-cavity-synthetic.md. In short: it works only from a calibration that still matches (image-space error up to about 1.5 px; beyond about 3 px the matcher finds nothing and the report says "not applicable"); consistency is restored in all 13 experiments (jump 1-4 µm, recall of the true cameras) while the error against the truth stays at 0.1-1.0 px depending on the reference camera and on what was wrong; glass distance and thickness are never recovered (the cameras absorb them), the fluid index only partly; the volume keeps its shape to 0.35 %; a rotation of the whole rig by 180 degrees (+z cameras to -z, glass vectors with them) changes nothing; and the matcher itself assumes the cameras look roughly along lab z (a rig turned 90 degrees finds no matches).

What the lab found

bench/calibration_lab builds a camera rig whose imaging is computed by an independent 3D ray tracer (openptv2.synthetic_rig: Snell's law through air | glass | water, with optional glass tilt, wedge, wrong thickness or index) and breaks one assumption at a time, with the truth known. See docs/plans/2026-11-26-calibration-lab.md for the tables. In short:

  • The library's flat-layer multimedia model agrees with the independent ray tracer to 0.001 px.
  • A standard plate calibration absorbs a wrong glass thickness (+1 to +4 mm), a 1 mm glass offset, a 0.3° tilt or wedge and a 0.01 higher fluid index into the other camera parameters: 3–9 µm of position error and no visible jump inside the plate volume. Even four times larger (1.2° wedge, 0.04 index) the error stays at 26–36 µm.
  • What does hurt is camera drift after the plate was recorded (hundreds of µm of jump from 0.1 mm / 0.3 mrad), and tracer self-calibration removes it (jump < 1 µm, leave-one-camera-out 1.8 → 0.008 px).
  • Free glass thickness / fluid index in the self-calibration are not identifiable at the level where they matter (fitted n3 moved 1.41 → 1.416 for a true 1.45) and the effect they would remove is below 40 µm; they are available (free_multimedia, free_glass) and off.
  • The blob detector is unbiased (≤ 0.002 px) and its pixel locking is below 0.03 px.